Enclomiphene citrate by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for Enclomiphene citrate, with a class minimum gradient of 15 min, the method is a live explanation.

70 recordsmedian 99.44 %spec 98.5 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik3030minHigh99.36+0.0099.36ADEQUATE
Medutest1325minUnit99.64+0.0899.56ADEQUATE
VendorInvestigate7 n=732minHigh98.91-0.1199.02ADEQUATE
Nordanalyt7 n=735minHigh99.41+0.0299.39ADEQUATE
Kestrel7 n=715minUnit99.66+0.2499.42ADEQUATE
PeptideMeter6 n=628minHigh99.87-0.0599.92ADEQUATE
97.698.098.498.899.399.7100.1JanoshikJanoshik: Q1 98.99, median 99.36, Q3 99.64, n=30n=30PeptideMeterPeptideMeter: Q1 99.54, median 99.87, Q3 99.93, n=6outlier 98.30n=6MedutestMedutest: Q1 98.63, median 99.64, Q3 99.93, n=13n=13VendorInvestigateVendorInvestigate: Q1 98.50, median 98.91, Q3 99.33, n=7n=7NordanalytNordanalyt: Q1 98.93, median 99.41, Q3 99.47, n=7n=7KestrelKestrel: Q1 99.42, median 99.66, Q3 99.93, n=7outlier 98.48n=7 Purity (% area)
Figure. Enclomiphene citrate purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.