AssayIndex

AX-2025-020421

KPV · 5 mg declared · Xi'an Mokemei Biotechnology Co., Ltd. · lot MKM-2504-N4868

FAILPURITY ONLY
Report ID
AX-2025-020421
Compound
KPV
Compound class
Tissue-repair and immunomodulatory peptides
Supplier
MKM — Xi'an Mokemei Biotechnology Co., Ltd.
Batch / lot
MKM-2504-N4868
Declared label claim
5mg
Laboratory
Medutest
Assay date
26 May 2025
Submitted
20 Jun 2025
Quarter
2025-Q2
Submitter
f.duarte
Sample provenance
Supplier-supplied
Section 1

Results

Measured against the index acceptance specification for the tissue-repair and immunomodulatory peptides class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 25 min gradient, 214 nm99.93% area≥ 99.0PASS
Total impuritiesRP-HPLC, sum of reported peaks0.07% area≤ 1.0PASS
Largest single impurityRP-HPLC — Pro-Val dipeptide (−128 Da)0.04% area≤ 1.00PASS
Identitynot ordered on this submissionnot determinedrequired for VERIFIED statusNOT TESTED
Total peptide contentby difference from the gravimetric mass balance81.45%w/wNOT TESTED
Target-compound contenttotal peptide content × purity81.39%w/wNOT TESTED
Gross fill (as weighed)five-place balance, under nitrogen5.47mg5 mg declaredNOT TESTED
Recovered target peptidegross fill × target content4.452mg5 mg claimNOT TESTED
Recovery against label claimrecovered ÷ claim89.0%90.0 – 110.0FAIL
Water contentKarl Fischer, coulometric, duplicate4.41%≤ 10.0PASS
Counter-ion (acetate)anion-exchange ion chromatography13.83%≤ 12.0FAIL
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.31%NOT TESTED
Endotoxinkinetic chromogenic LAL0.18EU/mg≤ 5.0PASS
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundWhite lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear, colourless, no visible particulatesclear, no visible particulatesPASS
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.052.5 – 8.0PASS
Retention time (observed)RP-HPLC, 25 min gradient4.44min4.44 ± 0.20 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.460.870.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²56,324≥ 20,000PASS
Resolution to nearest impurityvs Lys-Pro dipeptide (−99 Da)3.91no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

06508130171952526034 KPV4.44 min · 99.93 % 4.07 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 0481216202428 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 25 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 56,324, and skew follows the reported asymmetry factor of 0.87. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.04 % impurity against a 99.93 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1KPV — main peak4.444,856,90899.93
2Pro-Val dipeptide (−128 Da)4.071,9440.04-128.095
3Lys-Pro dipeptide (−99 Da)4.141,4580.03-99.068
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

ComponentDetermination%w/w of cake
WaterKarl Fischer4.41
Counter-ion (acetate)ion chromatography13.83
Residual solvent + inorganic residueestimated0.31
Total peptide contentby difference81.45
Total100.00 CLOSES

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.1099.93divergence -0.80 pp vs batch median
Total impurities (% area)0.900.07
Largest single impurity (% area)0.440.04
Water (%)not stated4.41
Counter-ion (%)not stated13.83
Peptide content (%w/w)not stated81.45
Endotoxin (EU/mg)not stated0.18
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Supplier-supplied
Evidentiary weight
Weaker evidence. Read with the caveat below.
Chain of custody
Supplier despatch → laboratory direct (courier, cold pack, 3 d) → assay. Submitter did not handle the vial.
Assay date
26 May 2025
Submitted to the index
20 Jun 2025
Laboratory
Medutest, Ljubljana, Slovenia
Method basis
25 min gradient · unit-resolution MS
Verification status
PURITY ONLY
Section 6

Annotations 4

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

peak_integrator
99.93 % is normalised area, so it is silent on how much peptide is in the vial. Content and water are the numbers that answer that, and both are on this record. Purity is the good news; content is the actionable news.
sole_source
Where this is the only record on a compound, the index says so on the compound page rather than reporting a median over one observation. Provenance is the second caveat there, not the first.
label_claim
The claim on the vial describes a mass. Whether that mass is peptide, peptide-plus-salt, or gross fill is the single most consequential ambiguity in this market, and 13.83 % acetate is how much it is worth here.
salt_exchange
Where the label states a peptide-content basis, the counter-ion is already excluded and the arithmetic is simple. Where it does not, this figure is how much of the difference it accounts for.
Section 7

Related records

Same batch — lot MKM-2504-N4868 n=12

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2025-020420MKM-2504-N4868Janoshik2025-Q299.665.0893.8Buyer-sourced0FAIL
AX-2025-020422MKM-2504-N4868Medutest2025-Q299.935.28Laboratory-retained0FAIL
AX-2025-020423MKM-2504-N4868Janoshik2025-Q299.435.57102.2Buyer-sourced0PASS
AX-2025-020424MKM-2504-N4868PeptideMeter2025-Q299.935.47Buyer-sourced0PASS
AX-2025-020425MKM-2504-N4868Janoshik2025-Q299.685.77105.7Buyer-sourced0FAIL
AX-2025-020426MKM-2504-N4868Medutest2025-Q299.935.6297.9Buyer-sourced1FAIL
AX-2025-020427MKM-2504-N4868Kestrel2025-Q299.935.1595.4Buyer-sourced0FAIL
AX-2025-020428MKM-2504-N4868Nordanalyt2025-Q299.936.09Buyer-sourced0FAIL
AX-2025-020429MKM-2504-N4868PeptideMeter2025-Q299.875.26Buyer-sourced0PASS
AX-2025-020430MKM-2504-N4868PeptideMeter2025-Q299.725.44Buyer-sourced0PASS
AX-2025-020431MKM-2504-N4868PeptideMeter2025-Q299.765.50Buyer-sourced0FAIL
Other recent KPV records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042742GGPepsGGPE-2511-K1569Janoshik2026-Q298.405.09Buyer-sourced0FAIL
AX-2026-042259TFCTFC-2508-M9624Janoshik2026-Q199.933.3889.4Buyer-sourced1FAIL
AX-2026-042078HYBHYB-2511-P7317Nordanalyt2026-Q199.593.9298.8Buyer-sourced0PASS
AX-2026-042262TFCTFC-2508-M9624Janoshik2026-Q199.893.3292.2Buyer-sourced0FAIL
AX-2026-042260TFCTFC-2508-M9624PeptideMeter2026-Q199.932.8999.2Buyer-sourced0FAIL
AX-2026-042079HYBHYB-2511-P7317Janoshik2026-Q199.685.6794.7Buyer-sourced0FAIL
AX-2026-042257TFCTFC-2508-M9624Janoshik2026-Q199.542.8195.6Buyer-sourced0FAIL
AX-2026-042261TFCTFC-2508-M9624Medutest2026-Q199.933.2097.1Buyer-sourced0FAIL
Other recent MKM records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042942CJC-1295 (no DAC)MKM-2603-V3346Janoshik2026-Q399.523.43Supplier-supplied0PASS
AX-2026-043019AdipotideMKM-2606-K5012Janoshik2026-Q397.615.77105.1Buyer-sourced0PASS
AX-2026-042939CJC-1295 (no DAC)MKM-2603-V3346PeptideMeter2026-Q399.443.39Buyer-sourced0PASS
AX-2026-043020AdipotideMKM-2606-K5012Medutest2026-Q397.735.74Buyer-sourced0PASS
AX-2026-043022AdipotideMKM-2606-K5012Kestrel2026-Q397.696.33106.9Buyer-sourced0PASS
AX-2026-043023AdipotideMKM-2606-K5012Medutest2026-Q397.485.56103.7Supplier-supplied0PASS
AX-2026-042945CJC-1295 (no DAC)MKM-2603-V3346PeptideMeter2026-Q399.393.54Laboratory-retained0PASS
AX-2026-042938CJC-1295 (no DAC)MKM-2603-V3346Nordanalyt2026-Q399.473.6795.0Buyer-sourced0PASS