Cross-section

BPC-157 by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for BPC-157, with a class minimum gradient of 15 min, the method is a live explanation.

510 recordsmedian 99.16 %spec 98.0 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik24630minHigh99.13+0.0099.13ADEQUATE
Medutest7525minUnit99.21+0.0899.13ADEQUATE
PeptideMeter7428minHigh99.06-0.0599.11ADEQUATE
Nordanalyt4835minHigh99.42+0.0299.40ADEQUATE
VendorInvestigate4732minHigh98.92-0.1199.03ADEQUATE
Kestrel2015minUnit99.62+0.2499.38ADEQUATE
97.097.598.098.699.199.6100.1JanoshikJanoshik: Q1 98.71, median 99.13, Q3 99.61, n=246n=246PeptideMeterPeptideMeter: Q1 98.66, median 99.06, Q3 99.68, n=74n=74MedutestMedutest: Q1 98.78, median 99.21, Q3 99.68, n=75n=75VendorInvestigateVendorInvestigate: Q1 98.55, median 98.92, Q3 99.58, n=47n=47NordanalytNordanalyt: Q1 98.94, median 99.42, Q3 99.87, n=48n=48KestrelKestrel: Q1 99.21, median 99.62, Q3 99.93, n=20n=20 Purity (% area)
Figure. BPC-157 purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.