The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for Adipotide, with a class minimum gradient of 15 min, the method is a live explanation.
| Laboratory | Reports | Gradient | MS resolution | Median purity % | Published bias (pp) | Bias-adjusted | Method suitability |
|---|---|---|---|---|---|---|---|
| Medutest | 7 n=7 | 25min | Unit | 97.73 | +0.08 | 97.65 | ADEQUATE |
| Janoshik | 6 n=6 | 30min | High | 98.68 | +0.00 | 98.68 | ADEQUATE |
| VendorInvestigate | 4 n=4 | 32min | High | 98.11 | -0.11 | 98.22 | ADEQUATE |
| Kestrel | 3 n=3 | 15min | Unit | 98.61 | +0.24 | 98.37 | ADEQUATE |
| PeptideMeter | 2 n=2 | 28min | High | 98.26, 98.69 | -0.05 | — | ADEQUATE |