AssayIndex

ERP-2401-D8257

Tirzepatide · 30 mg declared · Shanghai ERP Biotechnology Co., Ltd. · manufactured 19 Jan 2024

CHARACTERISED
Lot code
ERP-2401-D8257
Supplier
ERP — Shanghai ERP Biotechnology Co., Ltd.
Compound
Tirzepatide
Compound class
Incretin analogues and co-agonists
Declared label claim
30mg
Manufacture date
19 Jan 2024
Index quarter
2024-Q2
Independent assays
12
Laboratories involved
Janoshik, PeptideMeter, Medutest
Provenance mix
Buyer-sourced ×11 · Laboratory-retained ×1
Section 1

Reproducibility across independent tests

Every report below is an observation of the same physical material. What they disagree about is measurement; what they agree about is the lot. The size of the disagreement is compared against what the contributing laboratories' own repeatability figures predict.

Independent assays 12
Median purity 98.54% n=12
Spread (max − min) 0.52pp n=12
Standard deviation 0.155pp n=12
96.296.997.598.298.999.5100.2This lot (ERP-2401-D8257)This lot (ERP-2401-D8257): Q1 98.38, median 98.53, Q3 98.56, n=12n=12All TirzepatideAll Tirzepatide: Q1 98.05, median 98.55, Q3 99.01, n=750outlier 96.49outlier 96.55outlier 96.57n=750 Purity (% area)
Figure 1. This lot's independent purity results against the whole Tirzepatide record set in the index. The vertical line is the index acceptance specification for the incretin analogues and co-agonists class, 97.0 % area. A narrow box on the top row against a wide box on the bottom row is the ideal reading: it means the disagreement between lots is much larger than the disagreement within this one.
Section 2

Member reports 12

Every report the index holds on this lot. Each is a page carrying its own chromatogram, mass spectrum, mass balance, provenance and caveats.

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2024-004849ERP-2401-D8257Janoshik2024-Q298.354.1495.6Buyer-sourced0FAIL
AX-2024-004850ERP-2401-D8257Janoshik2024-Q298.424.67Buyer-sourced0PASS
AX-2024-004851ERP-2401-D8257Janoshik2024-Q298.624.1594.1Laboratory-retained0PASS
AX-2024-004852ERP-2401-D8257Janoshik2024-Q298.614.5688.8Buyer-sourced1MARGINAL
AX-2024-004853ERP-2401-D8257Janoshik2024-Q298.544.50Buyer-sourced0PASS
AX-2024-004854ERP-2401-D8257PeptideMeter2024-Q298.554.4597.0Buyer-sourced0PASS
AX-2024-004855ERP-2401-D8257Janoshik2024-Q298.534.1589.5Buyer-sourced1MARGINAL
AX-2024-004856ERP-2401-D8257PeptideMeter2024-Q298.764.15Buyer-sourced0PASS
AX-2024-004857ERP-2401-D8257Janoshik2024-Q298.394.14104.8Buyer-sourced1MARGINAL
AX-2024-004858ERP-2401-D8257PeptideMeter2024-Q298.543.6488.9Buyer-sourced1MARGINAL
AX-2024-004859ERP-2401-D8257Medutest2024-Q298.264.36Buyer-sourced0PASS
AX-2024-004860ERP-2401-D8257PeptideMeter2024-Q298.244.3194.6Buyer-sourced0PASS
Section 3

By laboratory

Where more than one laboratory assayed this lot, the difference between them is a paired observation of their relative bias — the same material, the same time window, different instruments. Those pairs are what the index's bias estimates are built from.

LaboratoryReportsPurity results (% area)MedianGradientPublished bias (pp)Bias-adjusted
Janoshik798.35, 98.42, 98.62, 98.61, 98.54, 98.53, 98.3998.5330min+0.0098.53
PeptideMeter498.55, 98.76, 98.54, 98.2498.5528min-0.0598.59
Medutest198.2698.2625min+0.0898.18

This lot contributes 3 paired comparisons to the inter-lab analysis.

Section 4

The supplier's certificate for this lot

One number the supplier wrote, beside the median of the numbers other people measured on the same physical material. That comparison is only available because the index groups by lot.

FigureSupplier COAThis reportIndex note
Purity (% area)98.6298.35divergence +0.08 pp vs batch median
Total impurities (% area)1.381.65
Largest single impurity (% area)0.990.62
Water (%)not stated4.14
Counter-ion (%)not stated5.64
Peptide content (%w/w)not stated89.86
Endotoxin (EU/mg)0.700.40
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Impurity fingerprint

The named impurities reported across this lot's assays, with how consistently each was seen. A fingerprint is what allows two lot numbers held by different companies to be identified as one bulk.

Named impurityReported inMedian % areaRange
[D-Aib13]-diastereomer12 / 120.610.52–0.76
des-Aib2 analogue12 / 120.370.23–0.65
C20 diacid mono-amide12 / 120.060.04–0.08
trifluoroacetyl adduct (+96 Da)12 / 120.030.02–0.03
truncated (3-39)6 / 120.570.51–0.64
tirzepatide dimer6 / 120.060.05–0.07
des-Gly deletion (−57 Da)6 / 120.180.16–0.22
Met(O) sulfoxide (+16 Da)6 / 120.020.01–0.02
Section 7

What this rollup does not establish