AssayIndex
Quality system

System suitability

The injections that establish a chromatographic system was fit to produce the result, and the parameters the index checks before accepting a purity figure.

method reference3 sections3 pharmacopoeial references

System suitability is the check that the instrument, column and method were performing acceptably at the moment the sample was injected. Without it, a result describes an unknown system. It is required by every pharmacopoeia and it is documented on almost no research-peptide certificate.

Parameters

ParameterDefinitionIndex requirement
Plate count NN = 5.54 (t/w½)²≥ 20,000 for the main peak
Asymmetry Aswidth at 10 % height, ratio of rear to front segment0.80 ≤ As ≤ 1.50
Tailing factor T(a + b) / 2a at 5 % heightT ≤ 2.0
Resolution RR = 2(t₂ − t₁) / (w₁ + w₂)computed and reported; no criterion applied
RepeatabilityRSD of peak area, replicate injections≤ 1.0 % (n = 5)
Signal-to-noise at LOQpeak height / peak-to-peak baseline noise≥ 10
Retention repeatabilityRSD of retention time≤ 1.0 %

Resolution is the one row in that table with no number beside it, and the omission is deliberate. Resolution to whichever named impurity happens to elute closest is a property of the compound's own impurity profile and of the gradient the laboratory chose — a compound whose nearest related substance is a diastereomer 0.2 min away can never reach the figure a compound with a clean 2-minute gap reaches, and penalising it for that would be scoring chemistry rather than chromatography. The index therefore computes R, prints it in the system-suitability block of every record that has one, and attaches no verdict to it. Rule AX-R12 was drafted with a resolution limb and that limb was withdrawn before it adjudicated anything; the rule is an asymmetry rule and nothing else. What the index does ask for where separation is poor is the integration convention, which is a documentation requirement rather than an acceptance criterion.

Sequence design

  1. Blank injection — diluent only. Establishes that no peak in the sample chromatogram originates in the diluent, the vial, or a carry-over from a previous injection.
  2. System-suitability solution — five replicate injections of a reference or of the sample itself. Establishes area and retention repeatability.
  3. Resolution solution — the sample spiked with a known related substance, or a degraded sample, where the pharmacopoeial monograph being followed sets a resolution criterion. The index sets none of its own.
  4. Sample injections, bracketed by a repeat of the system-suitability injection at the end of the sequence. Bracketing detects column degradation across a long run.

Worked plate count and resolution

Main peak:      t = 18.42 min,  w½ = 0.212 min

  N = 5.54 × (18.42 / 0.212)²
    = 5.54 × (86.89)²
    = 5.54 × 7 550
    = 41 830          ✓ above 20 000

Nearest impurity: t = 17.60 min,  w½ = 0.204 min

Convert half-height widths to baseline widths (w ≈ 1.70 × w½ for a Gaussian):

  w₁ = 0.361 min,  w₂ = 0.347 min

  R = 2 × (18.42 − 17.60) / (0.361 + 0.347)
    = 2 × 0.82 / 0.708
    = 2.32            ✓ well resolved

Same impurity on a 15-minute gradient (retention gap scales down):

  gap = 0.82 × 0.5 = 0.41 min, widths roughly halved to 0.181 / 0.174

  R = 2 × 0.41 / 0.355 = 2.31

— resolution is approximately preserved for a resolved pair.
The pairs that fail on a short gradient are those separated by
diastereomeric selectivity, which is not preserved.

Normative references

Cited because they are the documents the acceptance criteria above are taken from. The index applies them as written and states every deviation.

  • USP <621> — Chromatography, system suitability
  • Ph. Eur. 2.2.46 — Chromatographic separation techniques
  • ICH Q2(R2) — Validation of analytical procedures

Related methods

Purity by reverse-phase HPLC (% area)

The index reference method RG-2, what it measures, what it does not, and why a purity figure without a stated gradient is uninterpretable.

Chromatographic purity

Peak integration practice

Drop-line versus valley-to-valley, tangent skim, baseline placement, and the reporting threshold — where a purity figure is actually decided.

Chromatographic purity

Method validation to ICH Q2(R2)

The eight validation characteristics, what each one demonstrates, and which of them the reports in this index actually document.

Quality system